Table 2 – Percentage agreement (within 10 dB) between tablet audiometer and conventional audiometer at frequencies of 250 Hz, 500 Hz, 1000, 2000 Hz, 4000 Hz, and 8000 Hz

From: Adult validation of a self-administered tablet audiometer

 

250 Hz

500 Hz

1000 Hz

2000 Hz

4000 Hz

8000 Hz

mean

mean excluding AC250 Hz and AC8000 Hz

Percent Agreement

85.9

91.8

97.1

96.9

100.0

85.5

92.9

95.7

95% low

76.0

83.2

89.9

89.5

94.0

74.7

88.8

92.6

95% High

92.2

96.2

99.2

99.2

100.0

92.2

97.6

98.9