Table 3 – Percentage agreement (within 5 dB) between tablet audiometer and conventional audiometer at frequencies of 250 Hz, 500 Hz, 1000, 2000 Hz, 4000 Hz, and 8000 Hz

From: Adult validation of a self-administered tablet audiometer

 

250 Hz

500 Hz

1000 Hz

2000 Hz

4000 Hz

8000 Hz

mean

mean excluding 250 Hz and 8000 Hz

Percent Agreement

61.6

75.7

85.7

90.9

93.4

65.1

78.7

84.9

95% low

50.2

64.8

75.7

81.6

84.3

52.8

69.9

77.7

95% High

71.9

84.0

92.1

95.8

97.4

75.7

89.2

92.1