Table 4 – 2 × 2 contingency table comparison of tablet audiometry to conventional audiometry when hearing loss defined as at least one threshold of 40 dB HL or greater in at least one ear

From: Adult validation of a self-administered tablet audiometer

 

Traditional Audiometry

Tablet audiometry (SHOEBOX)

Normal hearing

Hearing loss

Normal hearing

13

1

Hearing loss

0

25