Navigation and non-navigation CT scan of the sinuses: comparison of the effective doses of radiation in children and adults

Abstract

Background

The advent of 3D navigation imaging has opened new borders to the endoscopic surgical approaches of naso-sinusal inflammatory and neoplastic disease. This technology has gained in popularity among otolaryngologists for endoscopic sinus and skull base surgeries in both adults and children. However, the increased tissue radiation required for data acquisition associated with 3D navigation protocols CT scans is a source of concern because of its potential health hazards. We aimed to compare the effective doses of radiation between 3D navigation protocols and standard protocols for sinus computed tomography (CT) scans for both the adult and pediatric population.

Methods

We performed a retrospective cohort study through electronic chart review of patients undergoing sinus CT scans (standard and 3D navigation protocols) from May 2019 to December 2019 using a Siemens Drive (VA62A) CT scanner. The effective dose of radiation was calculated in mSv for all exams. Average irradiation doses were compared using a Student’s T-Test or a Kruskall–Wallis test when appropriate.

Results

A total of 115 CT scans were selected for analysis, of which 47 were standard protocols and 68 were 3D navigation protocols CT scans. Among these, 31 exams were performed on children and 84 exams on adults. For the total population, mean effective dose in the non-navigation CT scans was 0.37 mSv (SD: 0.16, N = 47) and mean effective dose in the 3D navigation sinus CT group was 2.33 mSv (SD: 0.45, N = 68). The mean difference between the two groups was statistically significant 1.97 mSv (CI 95% − 2.1 to − 1.83; P < 0.0001). There was a sixfold increase in radiation with utilization of 3D navigation protocols. The ratio was identical when the pediatric as well as the adult subset of patients were analyzed.

Conclusion

In our center, utilization of 3D navigation sinus CT protocols significantly increases radiation exposure. Otolaryngologists should be aware of this significant increase and should attempt to decrease the radiation exposure of their patients by limiting unnecessary scan orders and by evaluating 3D acquisition protocols locally with radiation physicists.

Level of evidence: Level IV.

Graphical Abstract

Background

The advent of 3D navigation imaging has opened new borders to the endoscopic surgical approaches of naso-sinusal inflammatory and neoplastic disease [1,2,3,4,5,6,7,8]. The outstanding precision obtained from the navigation systems increases the safety level of the endoscopic procedures [9,10,11]. The use of navigation systems has recently gained in popularity among the pediatric population, particularly in the fields of neurosurgery and otolaryngology. The use of this technology has been described for several pediatric surgeries, including intracerebral tumor resections, skull base tumor resections, choanal atresia cures as well as endoscopic sinus surgeries. 3D navigation can be particularly useful in a very young population, for whom all the anatomical structures are smaller and in whom there is an increased risk of complications [5, 10]. However, the increased tissue radiation required for data acquisition associated with navigation (3D) CT scans is a source of concern in the medical literature because of its potential health hazards [12]. Radiation-induced cancers are certainly the most feared complications. This risk is notably greater in children, due to their higher radiosensitivity than adults and to a longer lifespan after radiation exposure [13,14,15]. Several strategies for reducing radiation doses have been studied. They include a modification in acquisition parameters, the use of iterative reconstructive techniques, the use of eye lens shielding and the adoption of cone beam technologies [16,17,18]. As stated by Hoxworth and Lal [17], the best way to decrease radiation exposure is to avoid unnecessary scanner examinations. To our knowledge, no study has quantified the difference of effective dose of radiation caused by the use of a 3D navigation acquisition protocol compared to a standard protocol specifically for CT scans of the sinuses, either in children or adult patients. The aim of our study is therefore to compare the effective radiation doses associated with CT scans of the sinuses with a 3D navigation protocol to those with a standard protocol in our tertiary referral center. The objectives of this study were threefold: (1) To survey the prescription habits of 3D navigation sinus CT in our center; (2) To compare the effective doses of radiation between navigation and non-navigation CT scans of the sinuses in the pediatric and adult population and (3) To compare the effective doses of radiation between children and adults for a similar navigation protocol.

ABC of ionizing radiation

Ionizing radiation includes gamma rays, X-rays and the higher ultraviolet part of the electromagnetic spectrum as opposed to the lower ultraviolet part of the spectrum, visible light, infrared, microwaves and radio waves which are examples of non-ionizing radiation. As previously stated, exposure to ionizing radiation has been shown to increase the incidence of neoplasia. There is evidence suggesting that radiation exposure in childhood is associated with a higher risk of developing various types of cancers later in life including leukemia, thyroid, skin, breast and brain cancer [19,20,21].

Three dosimetry quantities have been defined to measure radiation: the absorbed dose, the equivalent dose and the effective dose. The absorbed dose corresponds to the amount of energy deposited in a substance (e.g., human tissue). The absorbed dose is measured in a unit called the gray (Gy). A dose of one gray is equivalent to a unit of energy (joule) deposited in a kilogram of a substance [22]. The equivalent dose considers the damaging properties of the various types of radiation and is based on the absorbed dose by a tissue or an organ. This weighted absorbed quantity is expressed in a unit called the sievert (Sv). Finally, the effective dose is a measure of the total detriment or risk, due to exposure to ionizing radiation. The effective dose calculation combines the absorbed dose to the whole body, the relative harm level of the radiation as well as the specific sensitivity of each organ to ionizing radiation. If the exposure to different organs or tissues is not uniform, the concept of effective dose is used. The most significant dose quantity for patients is certainly the effective dose, as it allows for comparison and evaluation of long-term risk [23, 24]. The basic idea is to express the risk from the exposure to a single organ or tissue in terms of the equivalent risk from an exposure to the whole body. The unit of effective dose is the sievert. The three dosimetry quantities are protection quantities defined by The International Commission on Radiological Protection (ICRP) [25].

In 2009, the National Council on Radiation Protection and Measurements (NCRP) has published Report 160: Ionizing Radiation Exposure of the Population of the United States [26]. This report evaluated the doses to the U.S. population from all sources of ionizing radiation in 2006. The total effective dose from all sources was estimated at 6.2 mSv/year per individual. The average annual effective dose results from ubiquitous natural background (50%) and medical exposure of patients (48%), estimated to be approximately 3 mSv for each [26]. The main sources of ionizing radiation were natural background radon and thoron gaz (37%) and medical exposure with CT (24%) and nuclear medicine (12%) (Figs. 1, 2). Considering only the natural background dose, the added amounts of radiation over childhood and a lifetime period of 80 years are respectively 50 mSv and 250 mSv. The previous NCRP Report 93 (1987) [27] detailed the exposure of the US population during the early 1980s. From ubiquitous natural background radiation, the average individual effective dose was estimated to be similar. However, the average yearly dose of ionizing radiation from medical exposures was about 0.5 mSv per person in the USA during the 1980s; by 2006, it had increased to 3.0 mSv per person [14]. The increase is primarily due to the use of CT scans (responsible for 1/2 the 2006 medical dose), nuclear medicine (1/4 the dose mainly due to nuclear cardiology) and interventional fluoroscopy [26, 28].

Fig. 1
figure 1

Sources of ionizing radiation (NCRP Report 160, 2009)

Fig. 2
figure 2

Effective doses of radiation (mSv) from different sources (total 6.2 mSv/year per individual—NCRP Report 160 (2009))

Materials and methods

This study was conducted at the Centre Hospitalier Universitaire de Québec – Université Laval after approval from the Board of Ethics. From May 2019 to December 2019, we performed an electronic chart review of all patients who underwent both navigation and non-navigation CT scans of the sinuses. Patients could only be classified in one of the two groups (navigation vs non-navigation). For example, if a patient had had both standard and 3D navigation sinus CT, he was automatically classified in the navigation group. We collected data for both the pediatric and adult population. The scanner used for all CT exams was a Siemens Drive (Syngo CT VA62A) and data acquisition for the 3D navigation was based on the current recommendations from Stryker Co. (Table 1). Data relative to the CT parameters, such as kilovoltage (Kv), milliampere-seconds (mAs) and volumetric computed tomography dose index (CTDIv) were collected and analyzed by a medical radiation physicist (M.C.).

Table 1 Comparison of parameters of standard and navigation sinus CT scan protocols for 3 navigation systems

The CTDIv is the primary dose measurement concept in CT scanning. The CTDI represents the average absorbed dose in mGy [27]. It is calculated by the scanner based on the radiation output for the particular scan which would be absorbed by a polymethylmethacrylate (PMMA, e.g., acrylic or Lucite™) cylinder model. The head model is a 14-cm length cylinder with a 16 cm diameter. It is typically referred to as the head CTDI phantom. It is not a dose administered to a specific patient, but it is meant to be a comparison metric for different scans. The dose-linear product (DLP) is calculated by multiplying the CTDI-vol (which is obtained for a centimeter) by the number of centimeters scanned. Its result is presented in units of mGy-cm.

The effective radiation of each scan was calculated in millisieverts (mSv), based on dose-length product (DLP) reported by the CT scanner and using a conversion factor obtained from Table 3 of the Report No. 96 of the American Association of Physicists in Medicine [29]. The use of DLP to estimate the effective dose appears to be a reasonably robust method for estimating the effective dose. The values of Table 3 [29] are for adults with standard physique and pediatric patients of various ages over various body regions. Conversion factor for adult head and neck and pediatric patients assume the use of the head CT dose phantom (16 cm). The conversion factors used for calculations were 0.013 (birth), 0.0085 (1-year-old), 0.0057 (5-year-old), 0.0042 (10-year-old) and 0.0031 (adult) (Table 3, [29]), [30].

The rest of the data collection was performed by the other co-authors (J.L. and N.V.P). The indications and the specialty of the physician who ordered the CT scan were obtained for all patients. The number of CTs performed for the same episode of care as well as the presence of a surgical intervention after CT were collected for the 3D navigation protocol group. Student’s t-test was used to compare the navigation to the non-navigation pediatric and adult effective dose values. It was also used to compare the pediatric versus the adult effective dose results. The significance level was set at P ≤ 0.05. The Kruskal–Wallis test was used to confirm the results because of the small size of the cohort. All analyses were performed with SAS university 3.8 software.

Results

Population description

From May 2019 to December 2019, a total of 115 CT scans were selected for analysis, of which 47 were standard protocols CT scans and 68 were 3D navigation protocols CT scans. A total of 31 children and 84 adults were included in the analysis. In the pediatric cohort, 14 patients (45%, mean age = 10.9 years old) underwent standard protocol non-navigation sinus CT scans and 17 underwent 3D navigation protocol sinus CT scans (55%, mean age = 9.8 years old). Demographic data of both pediatric and adult cohorts can be found in Table 2. In Table 3, the list of indications for the scans and the prescribing specialties are shown for both the pediatric and adult cohort. In the children 3D navigation group, the majority of patients had had only one CT examination for the same indication (7/11). Four patients had a standard CT before the 3D navigation sinus CT scan. Of the 17 patients, 11 underwent surgical intervention after the navigation CT scan. In the other 6 patients, the CT scan did not yield indications for surgical intervention. In both children and adults, most of the navigation CT scans were prescribed by otolaryngologists.

Table 2 Demographic data
Table 3 Standard and navigation CT scans: indications and prescribing physicians

In our adult cohort, we found 33 (40%) non-navigation and 51 (60%) 3D navigation CT scans. The latter were prescribed exclusively by otolaryngologists. Of the 51 patients, only 25 underwent or had an indication for sinus surgery following 3D navigation CT scanning. In the other 26 patients imaging results did not reach criteria for sinus surgery. In the latter group, 21 patients had one or more sinus CT for the same indication. The majority were standard sinus CT protocols (14 patients) and less often, patients had another 3D navigation sinus CT scan (7 patients).

Comparison of irradiation doses

The scanner protocols for non-navigation and navigation scans were respectively: cut thickness 0.6 mm versus 1 mm; 120 kV versus 120 kV; 120 mAs versus 390 mAs. For the total population of effective radiation doses, the mean effective dose in the non-navigation CT group was 0.37 mSv (SD: 0.16, N = 47) and the mean effective dose in the 3D navigation sinus CT group was 2.33 mSv (SD: 0.45, N = 68). The mean difference between the two groups was statistically significant: − 1.97 mSv (CI 95% − 2.1 to − 1.83; P < 0.0001). Significant differences were also found when both pediatric and adult cohorts were compared separately. The mean difference of effective radiation dose was of 2.06 mSv (IC 95%: − 2.17 to − 1.96 mSv; P < 0.0001) for the adult cohort and was 1.68 mSv (IC 95%: − 2.07 to − 1.29 mSv; P < 0.0001) for the pediatric cohort. Effective radiation dose averages for the 3D navigation sinus CT scan protocols were compared between the pediatric population and the adult population. We also found a statistically significant mean difference of 0.41 mSv (IC 95%: − 0.64 to − 0.17; P = 0.03) between the two groups, with a higher level in adults. No difference was found between the 2 groups for the standard sinus protocol. A summary of the findings is presented in Table 4.

Table 4 Summary of results

Discussion

Head versus head and neck conversion factors

The difference between the two conversion factors (head vs head and neck) does not come from the size of the scans but from the type of tissues/organs involved. The ‘’Head’’ set of values include the brain, skin and bones. The ‘’Head and Neck’’ was used because the data acquisition for CT scans of the sinuses is done supine. The coronal cuts extend all the way down to the upper teeth to include the gums and the roots of the teeth. These cuts also include a significant portion of the parotid glands and parts of the submandibular glands. Of course, this setting normally includes the upper part of the esophagus, thyroid gland and some bone medulla. These tissues are not included in CT scans of the sinuses. The absolute value of effective radiation that we reported must be considered as the maximal possible dose. This has no impact on the ratio between the non-navigation and the navigation effective doses.

Average effective radiation ratio: standard versus navigation protocol

We observed a sixfold increase in the average effective radiation dose between standard protocol sinus CT and 3D navigation which was identical in both children and adults. In the navigation protocol, there also was a statistically significant 20% increase in the adult radiation dose when compared to the radiation dose of the pediatric cohort. To our knowledge, this study is the first to directly compare the effective radiation dose in mSv of standard versus 3D navigation protocol sinus CT scan in an institution, in both an adult and pediatric population. Even if an average effective dose of 2 mSv from a navigation scan remains a small dose and the associated risks for a single exam are negligible, it is of paramount importance to know the true difference in effective radiation doses between non-navigation and navigation sinus CT scans to better raise awareness among otolaryngologists on the irradiation associated with the 3D navigation sinus CT protocols. The younger age of the patient and the possibility of repeated radiological exams for sinusitis or other conditions should prompt the physician to be meticulous in his radiological prescriptions.

Average effective dose: importance of the mAs

The radiation doses in any specific radiological protocol is determined by the reference mAs. In our institution, for a standard CT scan of the sinus, the reference value was 60 mAs as opposed to the navigation protocol which was 390 mAs. This means that the scanner will set its exposition parameters around these reference values. The larger reference mAs values are associated with a higher definition of the images with less noise. These reference values can be changed according to the level of definition required for adequate navigation. There is a subjective aspect in this decision from the surgeon’s standpoint. Also, the nature of the disease such as polyps may require a higher reference value setting due to the required contrast with the bony structures.

The increases of radiation doses caused by 3D navigation CT protocols of any particular body region has been common knowledge for quite some time [31, 32]. In two different multicenter surveys done in 2009 [32] and 2011 [33], the authors reported respectively a tenfold and 18-fold variation in minimum and maximum doses of radiation for various sinus CT scan imaging protocols. Such variations may be explained by many factors such as the specific parameters and also, the type of scanner. Even though some centers use a unique acquisition protocol, many centers still have a specific 3D navigation protocol and a specific standard CT protocol for sinus CT scans [17].

In our study, 24% (4/17) of pediatric patients and 41% (21/51) of adult patients undergoing 3D navigation sinus CT had already had a sinus CT for the same episode of care. In addition, 3D navigation sinus CT confirmed a surgical indication or was utilized during a surgical procedure in 65% (11/17) of pediatric patients and 49% (25/51) of adult patients. The dilemma persists. On one hand, initial imaging with a 3D navigation protocol may not yield indication for surgery, which results in a non-necessary increase in irradiation for a patient. On the other hand, if one prescribes a standard sinus CT scan which reveals positive findings and criteria for endoscopic sinus surgery, the patient may need a subsequent 3D navigation protocol sinus CT scan with further radiation exposure. This raises a few questions: (1) Is there a clinical pattern that would have better predictability for a positive scan and surgical indication? (2) Is there a regional or seasonal variation in the significance of these patterns? and (3) Are these patterns the same for adults and children? Furthermore, patients with chronic diseases or neoplastic diseases may undergo repeated CT scans during a lifetime.

In our center, the aforementioned data has urged us to work on globally accepted dose-reduction strategies for sinus CT scanning. Among the possible paths, radiation reduction strategies for sinus CT scans have been studied and include adjusting specific scanner parameters [34,35,36], using iterative reconstruction techniques [37] or using cone-beam technologies [38,39,40]. We have used eye shielding in the past but our CT scans were inaccurate for navigation. For our specific scanner, X-CARE allows to reduce direct X-ray exposure for the most dose-sensitive body regions, e.g. the breasts, thyroid gland or eye lens through partial scanning. It protects these areas from direct X-ray exposure by lowering the tube current for a certain range of projections. It automatically adjusts the tube current for the remaining projections to prevent deteriorating image quality. Reduced sensitive-area exposure up to 40% was obtained without loss of image quality.

It seems likely that the most effective way to reduce radiation would be to eliminate unnecessary radiologic examinations. With that in mind, using a universal sinus CT protocol that fits both standard and 3D navigation needs may be a way to decrease radiation exposure. Hoxworth and al [17] validated a unique acquisition protocol for both standard and 3D navigation sinus CT scans in their center. In their study, 6187 sinus CT scans were performed, from which 596 were used for image-guided endoscopic sinus surgery. Intra-operative guidance with these CTs were all deemed technically adequate by surgeons. Unfortunately, radiation dose reduction per patient was not significant and translated into an effective dose reduction around 0.8 mSv. Furthermore, no decrease in radiation exposure per CT examination was obtained with their new protocol. Therefore, using a single protocol of this sort brings another kind of dilemma: even if the irradiation per person could be reduced by limiting the number of examinations, patients who do not require surgery would ultimately end up being exposed to higher levels of radiation. Another way to reduce radiation exposure could be to reserve the prescription 3D navigation CTs to physician with a certain area of expertise (OHNS, neurosurgeons, etc.). Judicious prescription of standard sinus CT protocols should also be advocated for primary care physicians as well as other specialists. This could play a role in improving the diagnostic yield and utility of these exams.

Navigation based surgery is a useful tool for the surgeon and has been widely used in sinus surgery [41, 42]. The American Academy of Otolaryngology – Head & Neck Surgery has published a position statement regarding the appropriate indications for computer-guided sinus surgery. These indications include: (1) revision surgery; (2) distorted anatomy (developmental, traumatic or postoperative); (3) extensive sino-nasal polyposis; (4) pathology of the frontal, posterior ethmoid and sphenoid sinuses; (5) disease involving the skull base, orbit, optic nerve or carotid artery; (6) CSF rhinorrhea or other skull base defects and (7) benign or malignant sino-nasal neoplasms [41]. A surgeon’s choice to use 3D image guided surgery technology may rely on these official indications as well as his preference and comfort. Although the use of this technology has demonstrated several advantages, it remains essential to consider the negative effects of increased irradiation on patients [40]. This is even more important for the pediatric population, in which the use of image-guided sinus surgeries is becoming increasingly frequent [5, 8, 10].

Our study has some limitations. First of all, our measurements were obtained from a single CT scanner and a limited number of patients were included. However, because of the considerable difference in irradiation, statistically significant differences were obtained even with our small population. Also, using a single CT device allowed us to have more comparable measurements between the standard and 3D navigation groups. Secondly, radiation doses from 3D navigation protocols can vary greatly. Our scanner and CT protocols are different from other centers. Furthermore, the different navigation system companies still provide different acquisition settings for both standard and 3D navigation sinus CT scans (Table 1).

Our study therefore emphasizes the importance of reviewing one’s own center practices. A patient-oriented approach is suggested including the referring physicians, general practitioners, pediatricians and the attending otolaryngologists to determine the exact indications for navigation and non-navigation sinus scanning and its proper timing. Regarding radiation doses, the technical requirements of the image guidance system must be met including the slice thickness, their overlapping and the minimum field of view. A medical physicist must study the other technical aspects such as the clinical software applications available on the CT scan to minimize irradiation. In cooperation with the physicians, he also must work on optimisation of the acquisition parameters with maintenance of the quality of the images and their accuracy for 3D navigation. A feedback process should be established to monitor each institution’s progress in radioprotection.

Conclusion

Navigation based surgery is a useful tool for the surgeon and is now widely used for both pediatric and adult patients. In our center on a single scanner, utilization of 3D navigation protocols for sinus CT imaging was associated with a sixfold increase over non-navigation protocols in effective radiation dose. Otolaryngologists should be aware of the possibility of a significant increase in radiation doses for navigation CT and should attempt to decrease the radiation exposure for their patients. A patient-oriented approach is suggested including a review of the prescription practices by referring physicians and otolaryngologists, a close monitoring of the radiological protocols and a feedback process to insure optimal radioprotection.

Availability of data and materials

We did not seek approval from the research ethics board to share data that was used in this retrospective cohort study. This will be requested if needed.

References

  1. Leonard S, Reiter A, Sinha A, Ishii M, Taylor RH, Hager GD. Image-based navigation for functional endoscopic sinus surgery using structure from motion. Proc SPIE Int Soc Opt Eng. 2016;9784:97840V.

    PubMed  PubMed Central  Google Scholar 

  2. Wagner W, Gaab MR, Schroeder HW, Tschiltschke W. Cranial neuronavigation in neurosurgery: assessment of usefulness in relation to type and site of pathology in 284 patients. Minim Invasive Neurosurg. 2000;43(3):124–31.

    Article  CAS  Google Scholar 

  3. Westendorff C, Dammann F, Reinert S, Hoffmann J. Computer-aided surgical treatment of bilateral choanal atresia. J Craniofac Surg. 2007;18(3):654–60.

    Article  Google Scholar 

  4. Willems PW, van der Sprenkel JW, Tulleken CA, Viergever MA, Taphoorn MJ. Neuronavigation and surgery of intracerebral tumours. J Neurol. 2006;253(9):1123–36.

    Article  CAS  Google Scholar 

  5. Bergeron M, Leclerc JE. Is image guidance accurate in children sinus surgery? Int J Pediatr Otorhinolaryngol. 2015;79(4):469–73.

    Article  Google Scholar 

  6. Rajasekaran S, Kanna PR, Shetty TA. Intra-operative computer navigation guided cervical pedicle screw insertion in thirty-three complex cervical spine deformities. J Craniovertebr Junction Spine. 2010;1(1):38–43.

    Article  CAS  Google Scholar 

  7. Lusk R. Computer-assisted functional endoscopic sinus surgery in children. Otolaryngol Clin N Am. 2005;38(3):505–13.

    Article  Google Scholar 

  8. Shah UK, Daniero JJ, Clary MS, Depietro JJ, Johnston DR. Low birth weight neonatal choanal atresia repair using image guidance. Int J Pediatr Otorhinolaryngol. 2011;75(10):1337–40.

    Article  Google Scholar 

  9. Al-Swiahb JN, Al Dousary SH. Computer-aided endoscopic sinus surgery: a retrospective comparative study. Ann Saudi Med. 2010;30(2):149–52.

    Article  Google Scholar 

  10. Postec F, Bossard D, Disant F, Froehlich P. Computer-assisted navigation system in pediatric intranasal surgery. Arch Otolaryngol Head Neck Surg. 2002;128(7):797–800.

    Article  Google Scholar 

  11. Labruzzo SV, Aygun N, Zinreich SJ. Imaging of the paranasal sinuses: mitigation, identification, and workup of functional endoscopic surgery complications. Otolaryngol Clin N Am. 2015;48(5):805–15.

    Article  Google Scholar 

  12. Canada's Choosing Wisely guidelines specifically relate to the issues of CT scans for sinus disease. https://journalotohns.biomedcentral.com/articles/https://doi.org/10.1186/s40463-020-00406-9.

  13. Frush DP, Donnelly LF, Rosen NS. Computed tomography and radiation risks: what pediatric health care providers should know. Pediatrics. 2003;112(4):951–7.

    Article  Google Scholar 

  14. Goodman TR, Mustafa A, Rowe E. Pediatric CT radiation exposure: where we were, and where we are now. Pediatr Radiol. 2019;49(4):469–78.

    Article  Google Scholar 

  15. Ohana O, Soffer S, Zimlichman E, Klang E. Overuse of CT and MRI in paediatric emergency departments. Br J Radiol. 2018;91(1085):20170434.

    Article  Google Scholar 

  16. Bang M, Choi SH, Park J, Kang BS, Kwon WJ, Lee TH, et al. Radiation dose reduction in paranasal sinus CT: with feasibility of iterative reconstruction technique. Otolaryngol Head Neck Surg. 2016;155(6):982–7.

    Article  Google Scholar 

  17. Hoxworth JM, Lal D. Universal sinus computed tomography protocol for diagnostic imaging and intraoperative navigation. Allergy Rhinol (Providence). 2015;6(3):146–50.

    Article  Google Scholar 

  18. Kong Q, Lu F, Gao Y, Qiao P, Shao M, Xie M. Wide-detector CT combined iterative reconstruction in pediatric low-dose scan of the paranasal sinus. J X-ray Sci Technol. 2019;27(4):731–8.

    Article  CAS  Google Scholar 

  19. Leuraud K, Richardson DB, Cardis E, Daniels RD, Gillies M, O’Hagan JA, et al. Ionising radiation and risk of death from leukaemia and lymphoma in radiation-monitored workers (INWORKS): an international cohort study. Lancet Haematol. 2015;2(7):e276–81.

    Article  Google Scholar 

  20. Miglioretti DL, Johnson E, Williams A, Greenlee RT, Weinmann S, Solberg LI, et al. The use of computed tomography in pediatrics and the associated radiation exposure and estimated cancer risk. JAMA Pediatr. 2013;167(8):700–7.

    Article  Google Scholar 

  21. Pearce MS, Salotti JA, Little MP, McHugh K, Lee C, Kim KP, et al. Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet. 2012;380(9840):499–505.

    Article  Google Scholar 

  22. Sinclair WK. The present system of quantities and units for radiation protection. Health Phys. 1996;70(6):781–6.

    Article  CAS  Google Scholar 

  23. Harrison J, Day P. Radiation doses and risks from internal emitters. J Radiol Prot. 2008;28(2):137–59.

    Article  Google Scholar 

  24. Dietze P, Jolley D, Cvetkovski S, Cantwell K, Jacobs I, Indig D. Characteristics of non-fatal opioid overdoses attended by ambulance services in Australia. Aust N Z J Public Health. 2004;28(6):569–75.

    Article  Google Scholar 

  25. Harrison JD, Streffer C. The ICRP protection quantities, equivalent and effective dose: their basis and application. Radiat Prot Dosimetry. 2007;127(1–4):12–8.

    Article  CAS  Google Scholar 

  26. Schauer DA, Linton OW. NCRP report No. 160, ionizing radiation exposure of the population of the United States, medical exposure—are we doing less with more, and is there a role for health physicists? Health Phys. 2009;97(1):1–5. https://doi.org/10.1097/01.HP.0000356672.44380.b7.

    Article  CAS  PubMed  Google Scholar 

  27. National Council on Radiation Protection and Measurements. Ionizing radiation exposure of the population of the United States. National Council on Radiation Protection and Measurements report no. 93. Bethesda, Md: National Council on Radiation Protection and Measurements, 1987.

  28. Mettler FA Jr, Thomadsen BR, Bhargavan M, Gilley DB, Gray JE, Lipoti JA, et al. Medical radiation exposure in the U.S. in 2006: preliminary results. Health Phys. 2008;95(5):502–7.

    Article  CAS  Google Scholar 

  29. AAPM Report No. 096 - The measurement, reporting, and management of radiation dose in CT. 2008. https://doi.org/10.37206/97. ISBN: 978-1-888340-73-0.

  30. Schell B, Bauer RW, Lehnert T, Kerl JM, Hambek M, May A, et al. Low-dose computed tomography of the paranasal sinus and facial skull using a high-pitch dual-source system–first clinical results. Eur Radiol. 2011;21(1):107–12.

    Article  Google Scholar 

  31. Urbanski W, Jurasz W, Wolanczyk M, Kulej M, Morasiewicz P, Dragan SL, et al. Increased radiation but no benefits in pedicle screw accuracy with navigation versus a freehand technique in scoliosis surgery. Clin Orthop Relat Res. 2018;476(5):1020–7.

    Article  Google Scholar 

  32. Nauer CB, Eichenberger A, Dubach P, Gralla J, Caversaccio M. CT radiation dose for computer-assisted endoscopic sinus surgery: dose survey and determination of dose-reduction limits. AJNR Am J Neuroradiol. 2009;30(3):617–22.

    Article  CAS  Google Scholar 

  33. Stewart AE, Stalp JT, Futerman C, Vaughan WC. Variability of radiation exposure with in-office sinus computed tomography examinations. Int Forum Allergy Rhinol. 2011;1:313–8.

    Article  Google Scholar 

  34. Lell MM, May MS, Brand M, Eller A, Buder T, Hofmann E, et al. Imaging the parasinus region with a third-generation dual-source CT and the effect of tin filtration on image quality and radiation dose. AJNR Am J Neuroradiol. 2015;36(7):1225–30.

    Article  CAS  Google Scholar 

  35. Sohaib SA, Peppercorn PD, Horrocks JA, Keene MH, Kenyon GS, Reznek RH. The effect of decreasing mAs on image quality and patient dose in sinus CT. Br J Radiol. 2001;74(878):157–61.

    Article  CAS  Google Scholar 

  36. Brem MH, Zamani AA, Riva R, Zou KH, Rumboldt Z, Hennig FF, et al. Multidetector CT of the paranasal sinus: potential for radiation dose reduction. Radiology. 2007;243(3):847–52.

    Article  Google Scholar 

  37. Hoxworth JM, Lal D, Fletcher GP, Patel AC, He M, Paden RG, et al. Radiation dose reduction in paranasal sinus CT using model-based iterative reconstruction. AJNR Am J Neuroradiol. 2014;35(4):644–9.

    Article  CAS  Google Scholar 

  38. Hodez C, Griffaton-Taillandier C, Bensimon I. Cone-beam imaging: applications in ENT. Eur Ann Otorhinolaryngol Head Neck Dis. 2011;128(2):65–78.

    Article  CAS  Google Scholar 

  39. Bulla S, Blanke P, Hassepass F, Krauss T, Winterer JT, Breunig C, et al. Reducing the radiation dose for low-dose CT of the paranasal sinuses using iterative reconstruction: feasibility and image quality. Eur J Radiol. 2012;81(9):2246–50.

    Article  Google Scholar 

  40. Schulz B, Beeres M, Bodelle B, Bauer R, Al-Butmeh F, Thalhammer A, et al. Performance of iterative image reconstruction in CT of the paranasal sinuses: a phantom study. AJNR Am J Neuroradiol. 2013;34(5):1072–6.

    Article  CAS  Google Scholar 

  41. Beswick DM, Ramakrishnan VR. The utility of image guidance in endoscopic sinus surgery: a narrative review. JAMA Otolaryngol Head Neck Surg. 2020;146(3):286–90.

    Article  Google Scholar 

  42. Schmale IL, Vandelaar LJ, Luong AU, Citardi MJ, Yao WC. Image-guided surgery and intraoperative imaging in rhinology: clinical update and current state of the art. Ear Nose Throat J. 2020. https://doi.org/10.1177/0145561320928202.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

The authors received no financial support for the research, authorship, and/or publication of this article.

Author information

Authors and Affiliations

Authors

Contributions

NV-P contributed to conception and design, acquisition of data, analysis, drafting, revision and final approval. JEL and MC contributed to conception and design, analysis, drafting, revision and final approval. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Noémie Villemure-Poliquin.

Ethics declarations

Ethical approval and consent to participate

The procedures were in accordance with the ethical standards of the CHU de Québec – Université Laval. Ethical approval from the research ethics board of our institution was obtained (request 2021–5252). Informed consent from patients was not deemed necessary as only clinical-administrative data and as no personal data were included in the manuscript. Publication approval was obtained from the research ethics board of our institution.

Consent for publication

Patient consent not available.

Competing interests

All of the authors (NVP, JEL, MC) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Research involving human participants and/or animals

Our study did not include experimentation on humans or animals. It is a retrospective study and therefore only clinical-administrative data were used.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Villemure-Poliquin, N., Chrétien, M. & Leclerc, J.E. Navigation and non-navigation CT scan of the sinuses: comparison of the effective doses of radiation in children and adults. J of Otolaryngol - Head & Neck Surg 50, 66 (2021). https://doi.org/10.1186/s40463-021-00541-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40463-021-00541-x

Keywords